| | Equipment features: 1. Double-layer insulation structure with air-cooled interlayer; 2. High purity Al2O3 fiber refractory insulation material, maximum energy saving; 3. The high purity alumina coating on the refractory furnace can extend the service life; 4. High purity quartz furnace plug helps you establish a custom temperature gradient in the tube. 5. Perfect overtemperature, short coupling, broken cooling water protection measures, safety! 6. Temperature hole and observation window are interchangeable (single option). |
Model | NBD-VAH1200-20TIF |
Electrical Spec | AC220V 4KW |
| System vacuum | 5-10Pa |
Tmax | ≦1100 ℃ |
Heating rate | ≤ 20 ℃/min |
Heating zone size | Ø190mm *200 |
Cavity septum | double-deck |
Temperature field distribution | | As shown in the figure: This is the temperature zone diagram of the equipment for reference only |
Control system | | 1. 15 temperature curves can be pre-stored to avoid the trouble caused by repeated Settings of different experimental processes; 2. The experiment process is more intuitive and the operation is more convenient; 3.NBD-101E embedded operating system English and Chinese interchangeable graphical interface, 7 inch true color touch screen input, intelligent man-machine dialogue mode, non-linear style temperature correction; 4. With overtemperature alarm, break even warning, break cooling water, leakage protection, etc. |
Temp Accuracy | +/- 1 ℃ |
Heating element | | Annular heating Mo doped Fe-Cr-Al alloy heater, the temperature field is uniform and reliable |
Sealing system | | | |
Vacuum degree: ≤10Pa (mechanical pump) |
Gas supply system and pressure measurement | | Mechanical pressure gauge with damping oil, and 1 small flow high-precision float flowmeter. |
Net weight | 20KG |
Cautions | 1. When the equipment is used, the pressure in the furnace tube shall not exceed 0.15MPa (absolute pressure) to prevent damage to the equipment caused by excessive pressure; 2. When used under vacuum, the operating temperature of the equipment shall not exceed 1000℃. 3. When the internal pressure of the gas supply cylinder is high and the gas is passed into the furnace tube, the pressure reducing valve must be installed on the cylinder. It is recommended to choose a small pressure reducing valve for test. 4. When the temperature of the furnace body is higher than 1000 ° C, the furnace tube can not be in a vacuum state, and the pressure in the furnace tube needs to be equivalent to atmospheric pressure and maintained at normal pressure to prevent damage to the chamber; 5. Long-term use temperature of high-purity quartz tube under atmospheric pressure ≦1150℃; 6. When heating and sintering, the exhaust valve should be opened to avoid chamber pressure increase when heating and heating, resulting in chamber rupture. If it is necessary to close the air valve to heat the sample, it is necessary to always pay attention to the indicator number of the pressure gauge. If the pressure of the barometer is within the indicated safety range, if the absolute pressure and reverse pressure are ≦0.05Mpa or the positive pressure ≧0.15Mpa, the exhaust valve must be opened immediately to prevent accidents (such as chamber rupture); 7. When this equipment is used for high-temperature sintering, it must be connected to the water cooler to operate, so as to avoid melting of the high-temperature sealing silicone ring. 8. Carefully sintering materials containing alkaline ions, alkaline ions will cause crystallization damage to the cavity. |
Service | 2-year limited warranty with lifetime support (warranty does not cover consumable parts such as quartz cavities and O-rings). |