Box Type Thermogravimetric Analyser (TGA)

Release time: 2024-07-10 03:41:12 Number of views: 126

Box Type
Equipment Description:
 

It is developed for analysing the weight loss by combustion and pyrolysis (TGA) of materials. It includes several parts, such as balance, heating furnace, programmed temperature control system, recording system, etc. It can observe the change of sample mass with temperature and time during the process of heating up, constant temperature or cooling down, and it is equipped with pyrolysis material condensation collection device. It is widely used in research and development, process optimisation and quality control in the fields of plastics, rubber, coatings, pharmaceuticals, catalysts, inorganic materials, metal materials and composites.

Edit Content

Configuration Details

System features:
1、Stable and reliable control, easy to operate.
2、Advanced PID self-learning fuzzy control, high temperature control accuracy, maintained at ± 1 ℃.
3、Furnace lining adopts high-purity alumina lightweight fibre material, better heat preservation, energy saving and consumption reduction.
4、Data storage function, can save the important parameters of sintering, up to 30 days long (8 hours a day boot).
5、Formulation function, can pre-store more than 100 formulas.
6、Multiple anti-dumbing function, so that customers can use it more safely.
7, networking function, through the RJ45 interface, the use of TCP / IP protocol, you can make the system connected to the host computer (the host computer needs to install the appropriate software).

Edit Content

Product Model

NBD-TGA1200-M20

Main Functions

1. In the process of temperature increase, constant temperature or temperature decrease, it can automatically record the change curve of sample quality with temperature and time;

Electrical specifications

AC220V 4.0KW

Tmax

1150 ℃

Temperature control accuracy

±1℃

Measuring element type

K-type thermocouple 420mm

Heating temperature area size

L200*H200*D300mm

Furnace size

L680*H980*D610mm

Load cell range

220g

Accuracy

0.1mg or less

Suspension Wire and Crucible Weight

Approximately 30g

Crucible Material and Dimensions

Aluminum/stainless steel φ60*30mm

Minimum Load

0~270g

Recommended Heating Rate

≤10℃/min

Control System

 

1. Load cell capacity: 220g.

2. Temperature profile settings: stores 15 profiles, each with up to 30 temperature-time segments.

3. Control method: a PLC automatically controls the drive mechanism, temperature modules, gas flow, vacuum generation, and weighing modules.

4. NBD-101EP embedded operating system with a bilingual graphical interface, 7-inch full-color touchscreen, interactive controls, real-time heating-power display, and nonlinear sample-temperature correction. Experiment reports are generated automatically and experiment data can be exported repeatedly.

5. Data interface: USB, supporting data export to a USB drive and online printing.

6. Data recording: automatically saves data at the configured time intervals.

7. Weighing module: a dedicated module provides improved accuracy and stability.

Weighing Balance

 

A reliable balance for simple, fast, efficient weighing. Capacity: 220g; readability: 0.1mg; USB and RS232 connections; external calibration; password protection; robust construction; 10 applications for accurate performance.

The electromagnetic force-compensation load cell includes overload protection for consistent results under intensive use.
Touchscreen Operable While Wearing Gloves
The color touchscreen provides an intuitive interface, simple menus, and clear icons. It can be operated while wearing cotton or rubber gloves.
A metal base, stainless steel weighing pan, and overload protection support reliable performance during intensive daily use.

Net Weight

90KG

Operating Precautions

  1. Power on the balance 1 hour in advance.
  2. Set the sintering process profile and ensure the crucible is installed.
  3. After 1 hour, zero the balance if required, then place the sample in the crucible.
  4. Start sintering and monitor changes in sample weight.
  5. After sintering, export the data for analysis.

Precautions:

1. Avoid strong airflow around the balance.

2. Avoid strong vibration around the balance.

3. The displayed weight must not exceed 4g before zeroing through the touchscreen; otherwise use the zero button on the balance.

Service and Support

1 year warranty with lifetime support (consumable components, such as furnace tubes and seals, are excluded).

DisclaimerProduct images, descriptions, and technical specifications are for reference only. Details may change or differ from the current equipment. Please confirm requirements and specifications by email. This information does not constitute an offer or commitment.Website information may be updated without prior notice.

Related Products

Edit Content
Click on the Edit Content button to edit/add the content.
Edit Content
Click on the Edit Content button to edit/add the content.

Recommended Products

Horizontal vacuum atmosphere furnace

Hydrogen reduction integrated sintering furnace (1500 ℃)

Steam generator for tube furnaces

Three temperature zone vertical tube furnace (fluidized bed FBCVD furnace)